Siemens MULTIX PRO

A pleasant atmosphere, comfort, and short examination times are becoming increasingly important in radiology departments. This is where our MULTIX systems really shine, especially with their operation and handling.
- Robust and easy-to-use floor-mounted system
- Selectable fixed table heights or height-adjustable table
- Outstanding image quality with minimal dose
- Simple, ergonomic, accurate patient-oriented operation
To meet customers' diverse application needs, there are various versions of the MULTIX PRO radiographic system.
- Compound patient tabletop of ± 0.6 Al; 240 cm long, 80 cm wide; longitudinal travel ± 48 cm, transverse travel ± 14 cm; table height 59.8 cm to 88.8 cm ± 0.5 cm, motorized adjustment.
- Tube support with a focus-floor distance from <=45 cm to 205 cm; travel range 160 cm, stand rotation ± 90; tube rotation ± 120° around horizontal axis.
Catapult Bucky and Cassette Programs
- All cassettes (metal or plastic) in compliance with DIN, IEC and ANSI: formats from 13 cm x 18 cm to 35 cm x 43 cm (5" x 7" to 14" x 17").
- Scatter radiation grid Pb 12/40; fo 115 cm.
- Travel range of catupult Bucky to the left 33 cm, to the right 28 cm.
- Tomographic exposures* at SID 115 cm or 102 cm at a digitally adjustable tomo height of 3 mm to 250 mm; tomo angle 40° with 1.2 s and 2.0 s; 30° with 0.8 s; 20° with 0.6 s; 8° with 0.4 s and 0.8 s.
- Tablefilm distance 53 mm
Imaging Components
- X-ray generator POLYDOROS LX 30 Lite (32 kW at 100 kV in compliance with IEC) or POLYDOROS LX 50 Lite (50 kW at 100 kV in compliance with IEC) or POLYDOROS LX 50 (50 kW at 100 kV in compliance with IEC) or POLYDOROS 80 (80 kW at 100 kV in compliance with IEC) or POLYDOROS SX 65 (65 kW at 100 kV in compliance with IEC) or POLYDOROS SX 80 (80 kW at 100 kW in compliance with IEC) as determined by the customer.
- OPTITOP 150/40/80 HC-100 X-ray tube for 150 kV, focus 0.6 and 1.0 (40/80 kW), or OPTILIX X-ray tube 150/30/50 HC-100) for 150 kV, focus 0.6 and 1.0 (30/50 kW).
- VERTIX PRO Bucky wall stand for right-/left-handed operation with vertical travel of 152 cm (center of cassette 38 cm to 190 cm ± 2 cm above floor) and all cassettes as with catapult Bucky; object film distance w 42 mm.